The Missouri Department of Health and Senior Services oversees the x-ray program for the state.
X-ray License in Missouri
The state of Missouri does not license personnel.
Registering an X-ray Device
All portable x-ray machines are required to be registered with Missouri Radiation Control Program (MRCP) upon installation and every 2 years.
Safety inspections are conducted at various periods depending on the facility and imaging modality type, ranging from 1-6 years. Inspections are conducted by DHSS staff or DHSS-approved Qualified Experts in radiation safety.
Regulations for Portable X-ray Equipment
19 CSR 20-10.160 Special Requirements for Mobile Medical Radiographic Installations
This rule establishes special requirements for mobile medical radiographic installations:
- All mobile equipment shall be provided with cones or frames so that the minimum target-to-skin distance is at least 18".
- The exposure control switch shall be so arranged that the operator can stand at least 6 feet from the patient and outside the primary beam for all exposures.
- A device shall be provided to terminate the exposure after a preset time.
Disclaimer: The information in this document is solely for reference material. Regulations change often. To be sure of the most up-to-date laws, contact your local and state agencies.